Gadolinium deposition in tissue following multiple contrast-enhanced MRI examinations

Othman Mohammed Mahzary

University of Leeds || UK

DOI PDF

Tab title
Gadolinium-based contrast agents (GBCAs) were considered extremely safe since their introduction for clinically used in 1988. However, in 2006, nephrogenic systemic fibrosis (NSF) was reported in patients with end-stage renal failure who were exposed to some GBCAs. In 2014, gadolinium deposition was found to occur in the brain tissues of patients exposed to multiple GBCA studies. This project aimed to review currently published studies regarding gadolinium accumulation in tissues after repeated GBCA administrations to identify and summarize the recent findings for this issue. Six scientific databases were electronically searched for relevant studies between 2009 and July 2016. The reference lists were also checked and tracked to find related articles. Studies were evaluated for relevance either by scanning their title and abstracts or reading the full text. The primary search yielded 1765 papers, which were then narrowed down to 14 studies that fulfilled the inclusion criteria. Two more studies were found by manually searching the reference lists. Of these, three studies assessed gadolinium retention in the brain tissues on animal models, while the remaining 13 were human research studies. Keywords: Gadolinium deposition, Multiple GBCAs, Dentate nucleus, Globus pallidus.

Introduction

Since the initial approval to use gadolinium-based contrast agents (GBCAs) by the competent authorities, around 450 million doses of GBCAs have been administered around the world (1). Compared with many other pharmaceuticals (i.e. iodinated contrast agents), GBCA has been extremely favorable and considered safe due to its small rates of adversative side effects (2). There is an association between using GBCA in patients suffering severe renal function and developing infrequent condition nephrogenic systemic fibrosis (NSF) where contractures in muscles, predominately skin or fibrotic changes may occur (3). Fortunately, the rapid development in the clinical practices of using GBCAs in renally impaired patients has fundamentally eliminated such clinical entity.

For a period of time, it had been widely thought that the ions of gadolinium remain chelated following GBCA intravenous administration; this was rapidly defeated. However, scientifically proven evidence has shown that the gadolinium traces remain in the brain, bones and other body organs in patients having a normal renal function (4). The tissue retention extent tends to be associated with the cumulative dosage. All linear and macrocyclic agents retain tissues of gadolinium where some of the existing data/studies suggest intraclass variability in such retention (5). However, the relationship between the exposure and retention of GBCA and their symptoms are still unclear due to the symptoms inconsistencies related to GBCA administration and dose threshold i.e. timing and location (6).

There are 9 of GBCAs that have been regulatory approved in the United States of America where each of them has its unique physical and chemical properties as shown in Table (1). However, linear GBCAs types have been recently removed from the European market because of the rising concerns about gadolinium retention (7). This difference is due to the difference in the country’s regulatory approach to using GBCAs. Besides, it could be due to the lack of scientific research on the adverse effects of those agents especially their bio distribution and their long-term retention in tissues. Consequently, conducting research studies on the safety of GBCAs agents to highlight their side effects is indispensable for the medical service providers and patients who need to undergo GBCA.

 

 

Table (1) GBCAs’ Physical and Chemical

Chemical Name Structure Ionicity Protein

Binding

Kobs (sec-1); T1/2 Log Ktherm Log Kcond Elimination

Half-Life

(min)

Injected Dose

Eliminated within 24

Hours (%)

Gadodiamide Linear Nonionic No 12.7; < 5 sec 16.9 14.9 77.8 ± 16 95.4 ± 5.5
Gadoversetamide* Linear Nonionic No 8.6; < 1 sec 16.6 15.0 NA NA
Gadopentetate dimeglumine Linear Ionic No 0.58; < 5 sec 22.5 18.4 96 ± 7.8 91 ± 13
Gadoxetate dimeglumine Linear Ionic Yes 0.16; < 4 sec 23.5 18.7 54.6–57 Amount remaining was too small to be detected
Gadobenate dimeglumine                Linear            Ionic            Yes                  0.41; < 5 sec                 22.6           18.4           70 ± 16    80–98

to 121 ± 36

Gadofosveset trisodium† Linear Ionic Yes 2.9 × 10-2; 24 sec 22.1 18.9            NA               NA
Gadoteridol Macrocyclic Nonionic No 2.6 × 10-4; 3.9 hr 23.8 17.1            94.2 ± 4.8 94.4 ± 4.8
Gadobutrol Macrocyclic Nonionic No 2.8 × 10-5; 43 hr 21.8 14.7            108             > 90

(72–393)

Gadoterate meglumine Macrocyclic Ionic No 2–8 × 10-6; 338 hr 25.6 19.3            84 ± 12 (F), 72.9 ± 17.0
120 ± 42  (F), 84.4 ±

(M)             9.7 (M)

Source: Mcdonald et al. (8)

 

Research objectives

The aim of the current research is to review some of the recent studies about gadolinium deposition in tissues following multiple administrations of GBCA to explore its side effects and impact on the patient.

Review of Included Studies

In this section, the researcher describes studies included in this research.

Animal research

Robert et al. (9) showed that the repeated administrations of the linear GBCA gadodiamide to healthy rats are associated with progressive and persistent T1 signal hyperintensity in the deep cerebellar nuclei (DCN), with Gd deposition in the cerebellum in contrast with the macrocyclic GBCA gadoterate meglumine for which no effect was observed. Jost et al. (10) evaluated, in rats, T1-weighted signal intensity in the deep cerebellar nuclei (CN) and globus pallidus (GP) up to 24 days after repeated administration of linear and macrocyclic gadolinium-based contrast agents (GBCAs) using homologous imaging and evaluation methods. They found increased signal intensity in the CN was found up to 24 days after multiple, extended doses of linear GBCAs. Robert et al. (11) evaluated Gd retention in the DCN of linear gadolinium-based contrast agents (GBCAs) compared with a macrocyclic contrast agent. The study revealed that repeated administrations of the linear GBCAs gadodiamide, gadobenate dimeglumine, and gadopentetate dimeglumine to healthy rats were associated with progressive and significant T1 signal hyperintensity in the DCN, along with Gd deposition in the cerebellum. This is in contrast with the macrocyclic GBCA gadoterate meglumine for which no effect was observed. Bussi et al. (12) explored the impact of single and cumulative doses of MultiHance on toxicity, pharmacokinetics, tissue gadolinium presence, behavior and neurological function in juvenile rats. Gadolinium presence was variable across tissues and decreased during the 60-day treatment-free period. The highest levels were noted in the femur and the lowest levels in the brain. Gadolinium presence in juvenile rat brain following single or repeated MultiHance administrations was minimal and non-impactful. Finally, Boyken et al. (13) determined the gadolinium (Gd) concentration in different brain areas in a pig cohort that received repeated administration of Gd-based contrast agents (GBCAs) at standard doses over several years, comparable with a clinical setting. The deduced that multiple exposures to macrocyclic gadobutrol are not associated with Gd deposition in brain tissue of healthy pigs.

Human research

Kanda et al. (14) revealed that High SI in the dentate nucleus and globus pallidus on unenhanced T1-weighted images may be a consequence of multiple previous gadolinium-based contrast material administrations. Errante et al. (15) found that the increase in the unenhanced T1 signal intensity has a linear relationship with the eMRIn in patients with MS and BM. Ramalho et al. (16) proved that gadolinium deposition occurs in the human brain after multiple gadolinium contrast administrations, despite an intact blood-brain barrier and normal renal function. Weberling et al. (17) found an increase in SI in the DN after serial injections of gadobenate dimeglumine. Quattrocchi et al. (18) found that the prolonged exposure of the capillary interface to dechelated gadolinium may contribute to the passage of the blood-brain barrier. Ramalho et al. (19) found an increased T1 signal change over time in patients who underwent gadobenate dimeglumine and had received prior gadodiamide compared to those without known exposure to previous gadodiamide. Adin et al. (20) showed that the repeated performance of gadolinium-enhanced studies likely contributes to a long-standing hyperintense appearance of dentate nuclei on precontrast T1-weighted-MR images. Kanda et al. (21) concluded that hyperintensity in the DN on unenhanced T1-weighted MR images is associated with previous administration of linear GBCA, while the previous administration of macrocyclic GBCAs showed no such association. Radbruch et al. (22) indicated that an SI increase in the DN and GP on T1-weighted images is caused by serial application of the linear GBCA gadopentetate dimeglumine but not by the macrocyclic GBCA gadoterate meglumine. In Cao et al. (23), unenhanced T1 signal hyperintensity was observed in the dentate nucleus after multiple administrations of gadopentetate dimeglumine, a linear ionic agent, but not after multiple administrations of gadobutrol, a macrocyclic GBCA. McDonald et al. (24) found that intravenous GBCA exposure is associated with neuronal tissue deposition in the setting of relatively normal renal function. Murata et al. (25) showed that the gadolinium deposition in the normal brain and bone tissue occurs with macrocyclic and linear protein interacting agents in patients with normal renal function. Finally, Stojanov et al. (26) found that patients with RRMS, SI within the dentate nucleus and globus pallidus increased on unenhanced T1-weighted images after multiple gadobutrol injections. Administration of the same total amount of gadobutrol over a shorter period caused greater SI increase.

Method and Procedures

The PubMed, Science Direct, ProQuest, Leeds library, Medline and Google Scholar databases were searched electronically for relevant studies that were published between 2009 and December 2018. Studies since 2009 were screened because some articles published before 2014 have reported hyperintensity in brain tissue on unenhanced T1-weighted images. The search terms used to find relevant articles are summarised in Table (2). The same search strategies were used for all the mentioned databases. The reference lists of the pertinent retrieved articles were manually tracked to identify additional relevant articles. The search process was completed on December 31, 2018.

Table (2) The search terms used to find relevant articles

Search Terms
Gadolinium MRI

AND

Deposit* OR retention* OR accumulated*

AND

Brain OR intracranial OR tissue.

The primary search process using the keywords mentioned above resulted in 3297 articles. After removing duplicates, 2985 papers remained. Of these, 2790 articles were deemed irrelevant by scanning the title and were excluded. An additional 154 papers were excluded by reading their abstract, which left 41 papers. A total of 22 of these articles were excluded after applying the previously mentioned exclusion criteria: 17 articles were reviews, six papers were commentary letters, and four were case reports. Of the remaining 27 studies, ten were excluded by reading the full text. These studies assessed hyperintensity in the DN on unenhanced T1W images due to several factors such as radiation therapy or progression of the existing disease but did not evaluate the impact of repeated GBCA administrations on unenhanced T1W SI. The reference lists of the relevant articles were manually searched, and two related articles (Adin et al., 2015; Weberling et al., 2015) were retrieved, increasing the number of relevant studies to 19.

The 19 included studies that fulfilled the inclusion criteria were initially classified based on study characteristics into animal research and human research. Five prospective studies (Robert et al., 2015; Jost et al., 2016; Robert et al., 2016, Bussi et al., 2017 and Boyken, et al., 2018) used animal models and compared the effect of multiple administrations of both macrocyclic and linear GBCAs on unenhanced T1W SI in deep cerebellar nuclei (DCN) involving the DN. The remaining 14 studies involved human patients.

The human studies were further divided into subgroups based on the type of GBCAs used and the study objectives. Seven studies (Kanda et al (14); Errante et al. (15); Ramalho et al (16); Weberling et al. (17); Quattrocchi et al (18); Adin et al. (20); Ramalho et al. (19) assessed the correlation between previous multiple linear GBCA administrations and gadolinium retention that caused hyperintense DN on unenhanced T1W images. McDonald et al. (24), Murata et al. (25), Bussi, et al. (17) and Lim et al. (27) examined the association between repeated GBCA administration and neural tissue deposition in deceased patients. Another three studies (Kanda et al. (21); Radbruch et al. (22); Cao et al. (23) compared the effects of previous multiple administrations of both macrocyclic and linear GBCAs on unenhanced T1W SI of the DN and GP. Stojanov et al. (26) evaluated the association between cumulative doses of macrocyclic GBCA, gadobutrol, and SI in both DN and GP on unenhanced T1W images. A summary of these studies is shown in Table (3).

Research Design and Sample

The following criteria were utilized to determine studies for the current review.

Inclusion criteria

  • Prospective or retrospective studies that were conducted either on human patients or animal models to evaluate gadolinium deposition in tissue.
  • Studies that assessed the correlation between increased SI in brain tissue on unenhanced T1W images and repeated administrations of different classes of GBCAs.
  • Studies performed to evaluate high SI in the brain tissue of patients with known diseases who received repeated administrations of GBCAs.

Table (3) lists the included studies.

Exclusion criteria

  • Studies that did not evaluate gadolinium deposition in tissue following repeated GBCA administrations.
  • Literature reviews and commentaries.
  • Systematic Reviews
  • Case reports.
  • Papers that reported general information about CE-MRI or chemical properties of GBCAs.

Data extraction

Data extracted from the relevant studies included the aim, study characteristics, type of GBCAs used in the study, interpretation of the results, and outcome of the studies.

 

Table (3) Summary of included studies

Study characteristics Type of GBCAs Methods of gadolinium deposition assessment
Study Study

 type

Population number Female Male Mean age

Years

Animal studies            
Robert et al. (9) Prospective 21 rats 21 0 Linear and macrocyclic —DCN/ cerebellar cortex SI ratio

—CP-MS to measure Gd concentration in brain tissue sample

Jost et al. (10) Prospective

 

60 rats 0 60 Linear and macrocyclic —DCN/pons SI ratio

—CSF spaces

Robert et al. (11) Prospective

 

40 rats 40 0 Linear and macrocyclic —DCN/ cerebellar cortex SI ratio

—ICP-MS for tissue analysis, R1 mapping

Bussi et al., (12) Prospective 486 rats 243 243 Linear and macrocyclic —ICP-MS to measure tissue samples of the brain (cerebellum, cerebral cortex, subcortical brain).
Boyken et al., (13) Prospective 13 pigs 0 13 Linear but Not Macrocyclic —ICP-MS to analyze tissues from multiple brain areas including cerebellar

and cerebral deep nuclei, cerebellar and cerebral cortex, and pons.

Human studies
Kanda et al. (14) Retrospective 19

16 (control)

9

10

10

6

67.7

73.5

Linear ionic —DN to pons and GP to thalamus SI ratios and Random coefficient model to assess SI
Errante et al. (15) Retrospective 38 MS**

37 BM**

27

18

11

19

47

62

Linear non-ionic —DN to pons SI ratio and relative change in the SI
Ramalho et al. (16) Retrospective 23 (linear non-ionic)

46 (linear ionic)

19

24

4

22

51.2

60.2

Linear ionic and non-ionic —DN/middle cerebellar peduncle (MCP) and GP/thalamus SI ratio and relative change in the SI were measured.
Weberling et al. (17) Retrospective

 

50

 

22

 

28

 

60.4

 

Linear ionic —DN to CSF and DN to pons SI ratio were measured
Quattrocchi et al. (18) Retrospective 46 40 6 66.5 Linear non-ionic —Calculation of DN to pons SI ratio
Ramalho et al. (19) Retrospective 18 (ionic & non-ionic)

44 (ionic)

16

23

2

21

56.5

60.4

Linear non-ionic —DN to MCP
Adin et al. (20) Retrospective 184 83 101 43.3 Linear —DN/MCP SI ratio —Assessed gadolinium retention among other confounders
Kanda et al. (21) Retrospective

 

127 69 58 60.4 Linear and macrocyclic —DN/cerebellum SI ratios

—Differences in SI between last and first GBCAs CE-MRI scan

Radbruch et al. (22) Retrospective

 

50 (macrocyclic)

50 (linear)

23

20

27

30

49.9

46.8

Linear and macrocyclic —DN/pons and GP/ thalamus mean SI ratios

—Regression analysis

Cao et al. (23) Retrospective 25 (macrocyclic)

25 (linear)

10

10

15

15

54

53

Linear and macrocyclic —DCP SI ratio before and after GBCA injections.
McDonald et al. (24) Retrospective

(Autopsy)

13 + (10 control) Linear —Neuronal tissue samples were taken from the brain and analyzed with ICP-MS
Murata et al. (25) Retrospective

(Autopsy)

9 + (9 control) 8 10 56.61 Linear protein binding GBCAs and macrocyclic —ICP-MS for gadolinium retention in tissue samples from bone, skin, and brain.
Stojanov et al. (26) Retrospective 58 37 21 42.38 Macrocyclic —DN/pons & GP/Thalamus SI ratios

—Correlation between No. of GBCA injections and high SI.

Lim et al., (27) Retrospective 44 15 29 54.3 Macrocyclic Evaluate the possibility of accelerated gadolinium accumulation in irradiated brain parenchyma

Notes: (*) Represents only the number of subjects included in the study who fulfilled the inclusion criteria. (MS**) patients with multiple sclerosis, (BM**) patients with brain metastasis. Gd= gadolinium, ICP-MS=inductively coupled plasma mass spectrometry.

Results and Discussion

The modern day MRI scanning can primarily be done in two methods, firstly without using contrast agents and the second, using contrast agents. MRI contrast involves using some extra component such as a dye for the scan, thus enabling better imaging in the process (28). The most commonly employed diagnostic agents or dyes make use of Gadolinium, which is a “rare earth metal” that acts as a “contrast enhancer,” thus making the areas of the body underscan “more visible” (29). MRI, as a “diagnostic imaging modality,” renders several benefits, as it is “non-invasive, delivers no radiation burden” and has excellent “spatial resolution” (31). This coupled with the increased visibility that contrast agents offer, MRI scans help in detecting what the exact problem a patient is suffering from and therefore enable appropriate diagnosis. In clinical imaging, basically the hydrogen atoms constituting water present in tissues are observed, thus the use of contrast agents “catalytically shortens the relaxation times of bulk water protons,” making tissues visible more accurately (31).

Some of the Gadolinium-based contrast agents used in MRI scans are “gadopentetate dimeglumine (Magnevist), gadoterate meglumine (Dotarem), gadoteridol (ProHance) gadodiamide (Omniscan), gadobenate dimeglumine (MultiHance), gadobutrol (Gadovist) and gadoversetamide (OptiMARK), Vasovist (gadofosveset)” etc (31). While some of these agents are ionic in nature, some are non-ionic and based on their thermodynamic property and kinetic stability different agents are used in different MRI scans. While the contrast agents seem highly relevant in making body tissues more visible, due to the chemical properties of the substances used in dyeing, people often seem to suffer from various side effects. Mild reactions to contrast agents include “nausea, headaches, dizziness, hives, rash, chills, pallor” etc, whereas moderate reactions include “tachycardia/ bradycardia, bronchospasm, wheezing, dyspnea, pronounced cutaneous reaction, laryngeal edema, pulmonary edema, hypertension” and so on. However, sometimes side effects can be highly wide-ranging and even cause life-threatening issues such as “cardiopulmonary arrest, clinically manifest arrhythmias, profound hypotension, unresponsiveness and convulsions” (32).

Adding to the issue of side effects caused by contrast based MRI scans are the high costs incurred in the process. The use of contrast agents and dyes undoubtedly leads to increased cost because gadolinium, as mentioned before, is a rarely found chemical substance. Furthermore, various alternatives can be used to perform scans instead of MRI, such as computed tomography (CT) scans, ultrasound, x-rays, digital subtraction angiography etc. In the case of scans for the brain, CT venography is a powerful tool, which facilitates the assessment of sinus thrombosis with cost-effectiveness (33).

Gadolinium as already discussed in the studies included in this paper. It is used as a contrast medium that makes certain tissues more vividly visible on an MRI scan. This agent is known to improve diagnostic accuracy in certain conditions such as inflammation as well as infectious diseases affecting bones, brain, spine and soft tissues. The contrast medium comprises of gadolinium ion bonded by a carrier molecule called cheating agent. The studies have shown that certain reported reactions that rarely occur are a brief headache, dizziness a few minutes following the injection. However, patients with reduced kidney function or kidney failure and hepatorenal syndrome should not be administered by this agent. Some patients were also found to be suffering from nephrogenic systemic fibrosis (NSF), a disease resulting in skin thickening and tightening as well as internal organ damage, occurring in people previously suffering from kidney abnormalities (34). People with normal kidney function are insusceptible to this disease.

Although the included studies showed an association between repeated administration of GBCAs and gadolinium retention in the brain tissue, the mechanism of gadolinium retention is still not clear. None of the included studies investigated whether the retained gadolinium in tissue remained chelated or in a free ion status. Some researchers have speculated on the form of retained gadolinium. Quattrocchi et al. (18) suggested that chelated GBCA could not pass through the intact blood-brain barrier, but it might dissociate at the DN capillary interface. However, McDonald et al. (24) demonstrated that gadolinium from injected GBCA could pass through the intact blood-brain barrier because the authors noted multiple punctate foci of gadolinium within the neural capillary endothelium, which was an indicator of the sound barrier. To dates, there have been no reports of clear mechanism or specific clinical significances of gadolinium deposition in tissues in patients with normal renal function.

The main limitation of the current review was that some of the included studies had a small sample size and were not excluded. In addition, the quality of the included studies was not assessed with a formal checklist, because the review aimed to identify evidence regarding gadolinium deposition in tissue.

Conclusion

In conclusion, this review revealed that there is some evidence to conclude that gadolinium retention in the tissue of patients with normal renal function is associated with multiple injections of less stable GBCAs. However, this evidence is still under investigation and unconfirmed by the US FDA. The reviewed studies’ results showed significant increases in the SI of the DN and GP on unenhanced T1W images after multiple administrations of linear GBCAs. Gadolinium retention in neuronal tissues is higher after successive administrations of linear non-ionic GBCAs than linear ionic GBCAs. Although most of the included studies reported no change in the SI of the DN and GP after multiple administrations of macrocyclic GBCAs, two studies reported an increase in the T1W SI in the DN following those administrations. Notably, these findings are unconfirmed due to methodological limitations, involving a small sample size. Gadolinium may accumulate in the bone, brain, and skin of patients with normal renal function; however, further large-scale studies are required to verify these findings. The mechanism of gadolinium deposition and whether the detected gadolinium in tissue remains chelated or as free ions remain unclear. Further studies are required to determine the specific type of GBCAs that can cause gadolinium retention and to identify the mechanism and clinical significance of this concern. Finally, throughout the reviewed studies, the clinical implications of gadolinium in tissue following multiple contrast-enhanced MRI examinations remain unclear and require further investigations. Besides, researchers and medical cadres are recommended to conduct more studies on the human being in order to find clear cut results concerning the retained gadolinium in tissue following multiple contrast-enhanced MRI examinations.

Acknowledgments

The author would like to thank Professor David Buckley for given his advice and time to support this project. The author would also like to thank Dr. Harry Tosumpas for giving the opportunity to do this project.

References

  • Mcdonald RJ, Levine D, Weinreb J, et al. Gadolinium Retention: A Research Roadmap from the 2018 NIH/ACR/RSNA Workshop on Gadolinium Chelates. Radiology. 2018;289(2):517-534.
  • Prince MR, Zhang H, Zou Z, Staron RB, Brill PW. The incidence of immediate gadolinium contrast media reactions. AJR Am J Roentgenol. 2011;196(2):W138-43.
  • Boyd AS, Zic JA, Abraham JL. Gadolinium deposition in nephrogenic fibrosing dermopathy. J Am Acad Dermatol. 2007;56(1):27-30.
  • Mcdonald RJ, Mcdonald JS, Kallmes DF, et al. Gadolinium Deposition in Human Brain Tissues after Contrast-enhanced MR Imaging in Adult Patients without Intracranial Abnormalities. Radiology. 2017;285(2):546-554.
  • Murata N, Gonzalez-cuyar LF, Murata K, et al. Macrocyclic and Other Non-Group 1 Gadolinium Contrast Agents Deposit Low Levels of Gadolinium in Brain and Bone Tissue: Preliminary Results From 9 Patients With Normal Renal Function. Invest Radiol. 2016;51(7):447-53.
  • S. Food and Drug Administration. FDA Briefing Document: Gadolinium Retention after Gadolinium-Based Contrast Magnetic Resonance Imaging in Patients with Normal Renal Function. https://www.fda.gov/downloads/AdvisoryCommittees/CommitteesMeetingMaterials/Drugs/MedicalImagingDrugsAdvisoryCommittee/UCM577014.pdf. September 8, 2017. Accessed January 9, 2019.
  • Tweedle M, Kanal E, Muller R. Considerations in the selection of a new gadolinium-based contrast agent. Appl Radiol 2014;(Suppl):1–11.
  • Mcdonald RJ, Levine D, Weinreb J, et al. Gadolinium Retention: A Research Roadmap from the 2018 NIH/ACR/RSNA Workshop on Gadolinium Chelates. Radiology. 2018;289(2):517-534.
  • Robert P, Lehericy S, Grand S, et al. T1-Weighted Hypersignal in the Deep Cerebellar Nuclei After Repeated Administrations of Gadolinium-Based Contrast Agents in Healthy Rats: Difference Between Linear and Macrocyclic Agents. Invest Radiol. 2015;50(8):473-80.
  • Jost G, Lenhard DC, Sieber MA, Lohrke J, Frenzel T, Pietsch H. Signal Increase on Unenhanced T1-Weighted Images in the Rat Brain After Repeated, Extended Doses of Gadolinium-Based Contrast Agents: Comparison of Linear and Macrocyclic Agents. Invest Radiol. 2016;51(2):83-9.
  • Robert P, Violas X, Grand S, et al. Linear Gadolinium-Based Contrast Agents Are Associated With Brain Gadolinium Retention in Healthy Rats. Invest Radiol. 2016;51(2):73-82.
  • Bussi S, Penard L, Bonafè R, et al. Non-clinical assessment of safety and gadolinium deposition after cumulative administration of gadobenate dimeglumine (MultiHance) to neonatal and juvenile rats. Regul Toxicol Pharmacol. 2018;92:268-277.
  • Boyken J, Frenzel T, Lohrke J, Jost G, Pietsch H. Gadolinium Accumulation in the Deep Cerebellar Nuclei and Globus Pallidus After Exposure to Linear but Not Macrocyclic Gadolinium-Based Contrast Agents in a Retrospective Pig Study With High Similarity to Clinical Conditions. Invest Radiol. 2018;53(5):278-285.
  • Kanda T, Ishii K, Kawaguchi H, Kitajima K, Takenaka D. High signal intensity in the dentate nucleus and globus pallidus on unenhanced T1-weighted MR images: relationship with an increasing cumulative dose of gadolinium-based contrast material. Radiology. 2014;270(3):834-41.
  • Errante Y, Cirimele V, Mallio CA, Di Lazzaro V, Zobel BB, Quattrocchi CC. Progressive increase of T1 signal intensity of the dentate nucleus on unenhanced magnetic resonance images is associated with cumulative doses of intravenously administered gadodiamide in patients with normal renal function, suggesting dechelation. Invest Radiol. 2014;49(10):685-90.
  • Ramalho J, Semelka RC, Ramalho M, Nunes RH, Alobaidy M, Castillo M. Gadolinium-Based Contrast Agent Accumulation and Toxicity: An Update. AJNR Am J Neuroradiol. 2016;37(7):1192-8.
  • Weberling LD, Kieslich PJ, Kickingereder P, et al. Increased Signal Intensity in the Dentate Nucleus on Unenhanced T1-Weighted Images After Gadobenate Dimeglumine Administration. Invest Radiol. 2015;50(11):743-8.
  • Quattrocchi CC, Mallio CA, Errante Y, et al. Gadodiamide and Dentate Nucleus T1 Hyperintensity in Patients With Meningioma Evaluated by Multiple Follow-Up Contrast-Enhanced Magnetic Resonance Examinations With No Systemic Interval Therapy. Invest Radiol. 2015;50(7):470-2.
  • Ramalho J, Semelka RC, Alobaidy M, Ramalho M, Nunes RH, Castillo M. Signal intensity change on unenhanced T1-weighted images in the dentate nucleus following gadobenate dimeglumine in patients with and without previous multiple administrations of gadodiamide. Eur Radiol. 2016;26(11):4080-4088.
  • Adin ME, Kleinberg L, Vaidya D, Zan E, Mirbagheri S, Yousem DM. Hyperintense Dentate Nuclei on T1-Weighted MRI: Relation to Repeat Gadolinium Administration. AJNR Am J Neuroradiol. 2015;36(10):1859-65.
  • Kanda T, Osawa M, Oba H, et al. High Signal Intensity in Dentate Nucleus on Unenhanced T1-weighted MR Images: Association with Linear versus Macrocyclic Gadolinium Chelate Administration. Radiology. 2015;275(3):803-9.
  • Radbruch A, Weberling LD, Kieslich PJ, et al. Gadolinium retention in the dentate nucleus and globus pallidus is dependent on the class of contrast agent. Radiology. 2015;275(3):783-91.
  • Cao Y, Huang DQ, Shih G, Prince MR. Signal Change in the Dentate Nucleus on T1-Weighted MR Images After Multiple Administrations of Gadopentetate Dimeglumine Versus Gadobutrol. AJR Am J Roentgenol. 2016;206(2):414-9.
  • Mcdonald RJ, Mcdonald JS, Kallmes DF, et al. Intracranial Gadolinium Deposition after Contrast-enhanced MR Imaging. Radiology. 2015;275(3):772-82.
  • Murata N, Gonzalez-cuyar LF, Murata K, et al. Macrocyclic and Other Non-Group 1 Gadolinium Contrast Agents Deposit Low Levels of Gadolinium in Brain and Bone Tissue: Preliminary Results From 9 Patients With Normal Renal Function. Invest Radiol. 2016;51(7):447-53.
  • Stojanov DA, Aracki-trenkic A, Vojinovic S, Benedetto-stojanov D, Ljubisavljevic S. Increasing signal intensity within the dentate nucleus and globus pallidus on unenhanced T1W magnetic resonance images in patients with relapsing-remitting multiple sclerosis: correlation with the cumulative dose of a macrocyclic gadolinium-based contrast agent, gadobutrol. Eur Radiol. 2016;26(3):807-15.
  • Lim WH, Choi SH, Yoo RE, et al. Does radiation therapy increase gadolinium accumulation in the brain?: Quantitative analysis of T1 shortening using R1 relaxometry in glioblastoma multiforme patients. PLoS ONE. 2018;13(2):e0192838.
  • Estelrich J, Sánchez-martín MJ, Busquets MA. Nanoparticles in magnetic resonance imaging: from simple to dual contrast agents. Int J Nanomedicine. 2015;10:1727-41.
  • Questions and Answers on the Review of Gadolinium-Containing Contrast Agents. EMEA. 2013. Retrieved January 9, 2019, from http://www.ema.europa.eu/docs/en_GB/document_library/Referrals_document/gadolinium_31/WC500015635.pdf
  • Caravan, P. Strategies for Increasing the Sensitivity of Gadolinium-Based MRI Contrast Agents. Chemical Society Reviews. 2006. Retrieved January 9, 2019, from http://clarke.physics.uwo.ca/journalclubpdf/jcarchives/gad_review_caravan.pdf
  • Ersoy H, Rybicki FJ. Biochemical safety profiles of gadolinium-based extracellular contrast agents and nephrogenic systemic fibrosis. J Magn Reson Imaging. 2007;26(5):1190-7.
  • Robbins, JB & Pozniak, MA. Contrast Media Tutorial. American College of Radiology. 2010. Retrieved January 1, 2019, from https://www.radiology.wisc.edu/fileShelf/contrastCorner/files/ContrastAgentsTutorial.pdf
  • Smith SM, Jenkinson M, Woolrich MW, et al. Advances in functional and structural MR image analysis and implementation as FSL. Neuroimage. 2004;23 Suppl 1:S208-19.
  • Cavallini M, Piano S, Romano A, et al. Terlipressin given by continuous intravenous infusion versus intravenous boluses in the treatment of hepatorenal syndrome: A randomized controlled study. Hepatology. 2016;63(3):983-92.

 

 

 

 

ترسب الجادولينيوم في الأنسجة بعد الخضوع المتعدد للتصوير بتقنية الرنين المغناطيسي المحسنة بالصبغة

Tab title
تعتبر عوامل التباين القائمة على الجادوليوم (GBCAs) آمنة للغاية منذ إدخالها للاستخدام في مجال التصوير في عام 1988. وعلى الرغم من ذلك، في العام 2006، تم اكتشاف أضراراً صحية للجهاز على المرضى الذين يعانون من الفشل الكلوي بعد تعرضهم لعدة مرات من التصوير. وفي العام 2004، تم اكتشاف بعض من ترسبات الجادوليوم في أنسجة دماغ بعض المرضى الذين تعرضوا للتصوير عدة مرات. في ضوء ذلك، هدفت الدراسة الحالية إلى استعراض ومراجعة بعض الدراسات المنشورة حاليًا فيما يتعلق بتراكم الجادولينيوم في الأنسجة بعد التعرض للتصوير عدة مرات، وذلك لتحديد وتلخيص النتائج الحديثة لهذه المسألة. ولتحقيق ذلك، تم البحث في ستة قواعد بيانات علمية إلكترونياً للدراسات ذات الصلة بين الأعوام 2009 وديسمبر 2019 حيث تم فحص قواعد البيانات ومراجعتها وتتبعها للعثور على المقالات ذات الصلة. تم تقييم الدراسات لملاءمتها عن طريق مسح عناوين وملخصات أو قراءة النص الكامل. شمل البحث المبدئي 1765 بحثًا ، ومن ثم تم حصرها في 14 دراسة حققت معايير الاختيار. تم الحصول على اثنين من الدراسات عن طريق البحث يدويا في قوائم المراجع. من هذه ، قيمت ثلاث دراسات احتباس الجادولينيوم في أنسجة المخ على نماذج حيوانية، في حين أن الثلاثة المتبقية كانت دراسات بحثية على البشر. الكلمات المفتاحية: ترسب الجادوليوم، التصوير المتعدد بالرنين المغناطيسي، النَّواةُ المُسَنَّنَة، الكُرَةُ الشَّاحِبَةُ الإِنْسِيَّة.